Cardiac monitoring with a LINQII tag

An OpenCollar Edge can follow one LINQII cardiac tag worn by the same animal. The tag advertises over Bluetooth, the collar listens for it and relays what it heard on port 15, and Protect stores the readings as ordinary measurements of the collar, which means of the animal the collar is on (decision D282, phase 34).

The firmware calls the tag a "cardiac monitoring device Q", CMDQ for short, and so do the field names.

What the tag sends, and what it means

One record per sighting. Four of the values can be read directly, and five cannot:

Reading Metric What it is
Heart rate heart_rate Beats per minute, derived from the median time between R peaks, which the tag sends in tens of milliseconds
Heart rate variability heart_rate_variability RMSSD in milliseconds, the square root of the tag's mean squared successive R-R difference. Firmware 6.9.0 and later only
Body temperature cmdq_temperature Degrees Celsius from the tag's raw value
A reading happened cmdq_success True when the tag's temperature came through, which is how the firmware's own decoder judges a reading
R-R median cmdq_rr_median The measured value the heart rate is derived from, in tens of milliseconds
Activity, average and highest cmdq_activity_average, cmdq_activity_max Numbers the tag sends. The manufacturer does not publish what they mean or in what unit. They are not the collar's own accelerometer activity
Active minutes in the last hour cmdq_active_min_in_last_hour Reads as minutes, though the unit is not published
Impedance cmdq_impedance A number the tag sends. Its meaning is not published; it reads like a contact quality, but that is a guess
Mode sum cmdq_rr_median_modesum A number the tag sends. Its meaning is not published
Raw values cmdq_raw_temperature, cmdq_hrv_raw What the temperature and the HRV are derived from, kept so the derivation can be checked

The five without a published meaning are stored as the numbers they are. Nothing in Protect interprets them, and the metric registry says so in each one's description. Their definition sits in a private issue of the firmware's upstream repository; when it is shared, the descriptions can be filled in without touching any stored data.

A sighting is not a reading

The tag can be heard and have nothing to say: rr_median is zero, the temperature is zero, and cmdq_success is false. That is normal, not a fault. It usually means the tag is not against the skin.

This matters more than it sounds, because a quiet tag and a calm animal look the same in a heart rate chart. Every figure Protect shows is accompanied by how many sightings there were and how many of them carried a reading. When no reading came through, no heart rate, no temperature and no variability are written at all, rather than zeros that would drag an average down. The raw fields are still stored, so the record is complete.

How often it reports

The collar does not hear the tag continuously. It scans on a schedule and reports in batches:

Setting What it does
cmdq_enabled Switches the scanner on. Switching it off while a scan runs stops the scan
cmdq_searched_mac_address The Bluetooth address of the tag this collar follows. One collar, one tag
cmdq_scan_duration How long a short scan listens, in milliseconds
cmdq_search_interval How long to wait between scans, in seconds. Set it to the tag's advertising interval
cmdq_on_no_detection_wait_duration How long to wait after a long scan found nothing, in seconds
cmdq_reporting_interval How often the detections held in the buffer become a port 15 message, in seconds
cmdq_report_zero_messages_to_be_sent Report an empty message when nothing was detected between two reporting intervals. A debugging aid: it proves the scanner is running

The firmware uses four kinds of scan (immediate, short, long and a timeout scan) to find the tag's rhythm and then stay in step with it. A LINQII advertises a burst every three minutes, so a reporting interval of several minutes usually carries a few detections per message.

These settings are in the device's settings catalogue, so they are read and set on the device's Settings tab like any other (ADR 0035). The command Send latest CMDQ results asks the collar to report what it has now.

Where the readings appear

  • On the live map, an animal's panel carries a Heart rate row with the newest reading; clicking it unfolds the trend over the period, the way the battery does.
  • On the Data tab of the entity or the device, in the metrics table and the Data explorer, under the physiology category.
  • In the Cardiac monitoring analysis module, which is where the figures that need a method live: the daily rhythm, a resting heart rate, and what the collar managed to hear.

Rules and automations can use the readings like any other measurement: a rule on heart_rate fires the same way one on battery_voltage does. Protect ships no such rule, because a healthy heart rate depends on the species, the age, the season and what the animal was doing a minute ago, and none of that is in the data.

Firmware differences

Firmware Record Note
4.x none No CMDQ module. A port 15 frame from such a device is noted on the trace and the delivery is kept
6.1 to 6.8 13 bytes No HRV
6.9.0 and later 15 bytes HRV added

The record length comes from the firmware version the device reports, never from the frame, so a device that sends an unexpected number of bytes leaves a note rather than eleven fields read at the wrong offsets.

Checking a decode

The firmware's own reference decoders are vendored under tests/fixtures/payloads/opencollar/decoders/ and the golden test runs every recorded frame through them, so what Protect reads cannot drift from what the firmware means. The full record layout is in OpenCollar protocol research, section 3.13, and the design of this phase is in docs/CARDIAC_MONITORING_PLAN.md.